{"id":23198,"date":"2023-01-31T12:38:27","date_gmt":"2023-01-31T11:38:27","guid":{"rendered":"https:\/\/genesys-offenburg.de\/?post_type=smart-docs&#038;p=23198"},"modified":"2026-04-02T15:29:19","modified_gmt":"2026-04-02T13:29:19","slug":"process-sequence","status":"publish","type":"smart-docs","link":"https:\/\/genesys-offenburg.de\/zh-hans\/support\/adma-manual\/measuring-procedure\/process-sequence\/","title":{"rendered":"Measurement process sequence"},"content":{"rendered":"\n<p>After powering up, the system runs through a boot process which ends in standby mode. The ADMA&#8217;s measuring mode can then be started via the configured data interface or operated with the Start and Stop commands.<\/p>\n\n\n\n<p>Measurements are performed in this specific sequence:<br>\u2022 Issue of a start command<br>\u2022 Receive the echo command from the ADMA<br>\u2022 Pre-alignment at a standstill for calibration and setting of start conditions<br>\u2022 Yaw angle initialization (Auto Init)<br>\u2022 Transient phase of the Kalman filter<br>\u2022 Continous measurement<br>\u2022 Issue of a stop command to terminate measurement. On issue of the stop command, the ADMA returns to the main menu and waits for new commands.<\/p>\n\n\n<div class=\"wp-block-image\">\n<figure class=\"aligncenter size-full\"><img decoding=\"async\" width=\"377\" height=\"850\" src=\"https:\/\/genesys-offenburg.de\/wp-content\/uploads\/2022\/08\/Bild_2022-08-15_114226646.png\" alt=\"\" class=\"wp-image-26718\" srcset=\"https:\/\/genesys-offenburg.de\/wp-content\/uploads\/2022\/08\/Bild_2022-08-15_114226646.png 377w, https:\/\/genesys-offenburg.de\/wp-content\/uploads\/2022\/08\/Bild_2022-08-15_114226646-133x300.png 133w, https:\/\/genesys-offenburg.de\/wp-content\/uploads\/2022\/08\/Bild_2022-08-15_114226646-300x676.png 300w\" sizes=\"(max-width: 377px) 100vw, 377px\" \/><figcaption class=\"wp-element-caption\">ADMA process sequence<\/figcaption><\/figure>\n<\/div>","protected":false},"excerpt":{"rendered":"<p>After powering up, the system runs through a boot proce [&hellip;]<\/p>\n","protected":false},"author":148,"featured_media":30266,"template":"","meta":{"_acf_changed":false,"_links_to":"","_links_to_type":0,"footnotes":""},"smartdocs_category":[809,848],"class_list":{"0":"post-23198","1":"smart-docs","2":"type-smart-docs","3":"status-publish","4":"has-post-thumbnail","6":"smartdocs_category-adma-manual","7":"smartdocs_category-measuring-procedure"},"acf":[],"_links":{"self":[{"href":"https:\/\/genesys-offenburg.de\/zh-hans\/wp-json\/wp\/v2\/smart-docs\/23198","targetHints":{"allow":["GET"]}}],"collection":[{"href":"https:\/\/genesys-offenburg.de\/zh-hans\/wp-json\/wp\/v2\/smart-docs"}],"about":[{"href":"https:\/\/genesys-offenburg.de\/zh-hans\/wp-json\/wp\/v2\/types\/smart-docs"}],"author":[{"embeddable":true,"href":"https:\/\/genesys-offenburg.de\/zh-hans\/wp-json\/wp\/v2\/users\/148"}],"version-history":[{"count":1,"href":"https:\/\/genesys-offenburg.de\/zh-hans\/wp-json\/wp\/v2\/smart-docs\/23198\/revisions"}],"predecessor-version":[{"id":48445,"href":"https:\/\/genesys-offenburg.de\/zh-hans\/wp-json\/wp\/v2\/smart-docs\/23198\/revisions\/48445"}],"wp:featuredmedia":[{"embeddable":true,"href":"https:\/\/genesys-offenburg.de\/zh-hans\/wp-json\/wp\/v2\/media\/30266"}],"wp:attachment":[{"href":"https:\/\/genesys-offenburg.de\/zh-hans\/wp-json\/wp\/v2\/media?parent=23198"}],"wp:term":[{"taxonomy":"smartdocs_category","embeddable":true,"href":"https:\/\/genesys-offenburg.de\/zh-hans\/wp-json\/wp\/v2\/smartdocs_category?post=23198"}],"curies":[{"name":"wp","href":"https:\/\/api.w.org\/{rel}","templated":true}]}}